Liquid chromatograph
The stable flow rate ensures the repeatability of qualitative and quantitative indicators of liquid chromatography.
Excellent wavelength repeatability ensures the consistency of energy and the quantitative index of liquid chromatography.
The adoption of filter group (6 pieces) structure reduces the influence of stray light on the analysis results, and improves the minimum detection concentration and linear range of the instrument.
Powerful and completely anti control workstation software.
Index of high pressure infusion pump
Settable flow range:
0.001 ml/min~ 10.000ml/min in increments of 0.001ml/min
Maximum operating pressure: 40MPa, upper and lower limits can be set, and automatic alarm can be given
Pressure fluctuation: <1% (1 ml/min, water)
Flow repeatability: ≤ 0.075%rsd (based on retention time)
Flow accuracy: ± 1%
Accuracy of binary high voltage gradient: ≤ 1%
Accuracy of binary high voltage gradient: ≤ 0.2%
Indicators of UV detector
Light source: deuterium lamp, hook lamp
Wavelength range: 190nm~800nm
Spectral bandwidth: 6nm
Wavelength accuracy: ± 1nm (deuterium lamp)
Wavelength repeatability: ≤ 0.2nm (deuterium lamp)
Linear range: ≤ 104
Noise: ± 0.75x10-5 Au (empty cell, 254nm, integration time 1s)
Drift: ≤ 1x10-4au/h (empty cell, 254nm, integration time 1s)
Minimum detection concentration: <5x10-9g/ml (naphthalene / methanol solution)
Flow cell volume: 10 μ L